Page 80
HydroRanger 200 – INSTRUCTION MANUAL
7ML19985FC03
mmmm
m
Open Channel
Monit
oring
6 Relay Model
Cut Throat Flume
Application Information
•
Similar to Parshall flume except that the floor is flat bottomed and throat has no
virtual length.
•
Refer to manufacturer’s specifications for flow equation and point of head
measurement.
Flume Size
(pipe diameter in inches)
Point of Measurement
cm
inches
4-12
2.5
1
15
3.2
1¼
18
4.4
1¾
21
5.1
2
24
6.4
2½
30
7.6
3
42
8.9
3½
48
10.2
4
54
11.4
4½
60
12.7
5
66
14.0
5½
72
15.2
6
Parameter
Index
Value
P600
G
1–Cut Throat Flume
P601
G
1.55
P603
G
Maximum Head (preset P007)
P604
G
Maximum Flow
P606
G
Time Units
Plan View
Summary of Contents for HydroRanger 200
Page 1: ...Instruction Manual July 2009 200 hydroranger ...
Page 232: ...Page 222 HydroRanger 200 INSTRUCTION MANUAL 7ML19985FC03 mmmmm Parameters Notes ...
Page 264: ...Page 254 HydroRanger 200 INSTRUCTION MANUAL 7ML19985FC03 mmmmm Appendix H Software Rev Notes ...
Page 278: ...Notes ...
Page 279: ......