HydroRanger 200 – INSTRUCTION MANUAL
Page 25
m
mmmm
Oper
a
tion
Setting the Default Auxiliary Reading
To maintain a constant variable display in the auxiliary reading area, set the default.
Example:
To leave the level reading on the screen and view the echo confidence in the auxiliary
reading field, set the following parameter:
Setting a Specific Auxiliary Reading
To display a second auxiliary reading, press
in RUN mode.
Example: To set
to display current temperature, go to P731:
Multiple Readings [6 relay model]
During differential or average operation (P001 = 4 /5), the display scrolls sequentially
through Point Numbers 1, 2, and 3. Point 3 is the difference between (or average of) Points
1 and 2.
Changing Number Scrolling Speed
See
Parameter Indexing
on page 35. All the instructions in the following procedures apply
to the hand programmer and assume that the HydroRanger 200 is activated.
Parameter
Index
Value
Description
P730
G
805
Auxiliary field defaults to P805
Parameter
Index
Value
Description
P731
G
912
Shows P912–Transducer Temperature
Parameter
Index
Value
Description
P732
G
5
Hold each value for 5 seconds
P
P