Reading Measurements by Operation
P921 Material Measurement
The distance in Units (P005) or % of Span (P007) between Empty (P006) and the
monitored surface.
P922 Space Measurement
The distance between the monitored surface and Span (P007).
P923 Distance Measurement
The distance between the monitored surface and the transducer face.
P001
P050 = 0
P050
≠
0
0 – OFF
- - - -
- - - -
1 – Level
P921
P924
2 – Space
P922
100% - P924
3 – Distance
P927
P927
4 – Difference [6 relay model] P921 (indexed)
P921 (indexed)
5 – Average [6 relay model]
P921 (indexed)
P921 (indexed)
6 – OCM [6 relay model]
P925
P925
7 – Pump Totalizer [6 relay
model]
P924
P924
Primary Index
Level
Values
Range: -999 to 9999 (view only)
Related
•
P005 Units
•
P006 Empty
•
P007 Span
Primary Index
Transducer
Values
Range: 0.000 to 9999 (view only)
Related
•
P007 Span
Primary Index
Transducer
Values
Range: 0.000 to 9999 (view only)