Remote Control Commands
R&S
®
Scope Rider RTH
410
User Manual 1326.1578.02 ─ 12
Return values:
<ResultActual>
Measured value
Range:
-100E+24 to 100E+24
Increment: 0
*RST:
0
Default unit: Hz
Usage:
Query only
COUNter<m>:INITiate
Resets the hardware and all statistical values, performs a single measurement and
ends in hold mode.
Suffix:
<m>
.
1..2
Usage:
Event
COUNter<m>:MEASure:FREQuency?
<Range>
This command sets all measurement parameters and trigger parameters to default val-
ues for frequency measurements, then immediately triggers a counter measurement
and returns the result.
This command corresponds to the commands
Suffix:
<m>
.
1..2
Parameters:
<Range>
<numeric value> | MIN | MAX | DEF
Defines the measurement range of the input signal for the coun-
ter. If probes are connected, the MIN | MAX | DEF values are
adapted accordingly.
<numeric value>
Power range in V. If necessary, the next higher range is
selected.
MIN
100 mV
MAX
300 V
DEF
100 mV
Return values:
<Counter result>
Measured frequency
Usage:
Query only
Counter Mode (R&S
RTH-K33)