Documenting Results
R&S
®
Scope Rider RTH
268
User Manual 1326.1578.02 ─ 12
12.3
Instrument Settings
To repeat measurements or tests at different times or perform similar measurements
with different test data, you can save the used configuration settings for later use. Fur-
thermore, it can be helpful to refer to the configuration settings of a particular measure-
ment when analyzing the results. Therefore, you can easily save the complete mea-
surement configuration including the display settings.
The R&S
RTH provides two ways to store the measurement configuration:
●
Save as fast setting
●
Save as saveset
12.3.1
Using Fast Settings
The R&S
RTH has 8 slots to save and load often used configuration settings very
quickly.
To save the current setting as fast setting
1. Press the [SHIFT] key.
2. Tap and hold one of the storage slots F1...F8.
The online keyboard opens.
3. Type the settings name and tap [
↵
].
The configuration settings are saved to the slot.
To load a fast setting
1. Press the [SHIFT] key.
2. Tap the storage slot that holds the required configuration.
The settings are loaded.
Alternatively, you can save and load fast settings in the "File" menu: [FILE] > "Settings"
> "Fast Settings".
Instrument Settings