Remote Control Commands
R&S
®
Scope Rider RTH
329
User Manual 1326.1578.02 ─ 12
POSitive
Rising edge
NEGative
Falling edge
EITHer
Rising and falling edges are considered (double data rate).
At double data rate, the edge at which the first bit of the pattern
is sampled is defined by
*RST:
POSitive
TRIGger:SPATtern:CSOurce:FIRStedge
<FirstClockEdge>
Sets the edge at which the first bit of the pattern is sampled.
The command is required if
is set to
Either
(double data rate).
Parameters:
<FirstClockEdge>
POSitive | NEGative | EITHer
*RST:
POSitive
TRIGger:SPATtern:ORDer
<BitOrder>
Defines if the data words start with MSBF (most significant bit first) or LSBF (least sig-
nificant bit first).
Parameters:
<BitOrder>
LSBF | MSBF
*RST:
MSBF
TRIGger:SPATtern:PATTern
<Pattern>
Defines the serial pattern to be triggered on.
Parameters:
<Pattern>
String that contains the pattern in binary format. The parameter
accepts the bit value X (don't care).
Example:
TRIGger:SPATtern:PATTern '11001100'
TRIGger:SPATtern:PATTern '110011XX'
15.3.5.13
Timeout Trigger (R&S
RTH-K19)
............................................................................................... 330
Waveform Setup