Remote Control Commands
R&S
®
Scope Rider RTH
387
User Manual 1326.1578.02 ─ 12
Parameters:
<State>
OFF | ON
*RST:
OFF
METer<m>:SENSe:NULL:STATe
<SetRelative>
METer<m>:SENSe:RELative:STATe
<SetRelative>
Enables or disables the relative measurement for the currently active measurement
type.
Relative measurements are not available for diode, continuity, and frequency measure-
ments.
Suffix:
<m>
.
1..4
R&S
RTH1002: always 1, omit the suffix
R&S
RTH1004: Selects the voltmeter.
Parameters:
<SetRelative>
ON | OFF
ON = 1, OFF = 0
*RST:
OFF
METer:SENSe:CAPacitance:NULL:VALue
<ReferenceValue>
Sets the reference value for capacitance measurements. The measurement result is
the difference of the measured sample and the reference value.
The value takes effect if relative measurement is on, see
.
Parameters:
<ReferenceValue>
Range:
+/- (1.1 * measurement range)
*RST:
0
Default unit: F
METer:SENSe:CURRent:AC:NULL:VALue
<ReferenceValue>
METer:SENSe:CURRent:DC:NULL:VALue
<ReferenceValue>
METer:SENSe:VOLTage:AC:NULL:VALue
<ReferenceValue>
METer:SENSe:VOLTage:DC:NULL:VALue
<ReferenceValue>
Sets the reference value for relative measurements. The measurement result is the dif-
ference of the measured sample and the reference value.
The value takes effect if relative measurement is on, see
.
Parameters:
<ReferenceValue>
Range:
+/- (1.1 * measurement range)
*RST:
0
Default unit: V (VOLTage) | A (CURRent)
Digital Multimeter (R&S
RTH1002)