Remote Control Commands
R&S
®
Scope Rider RTH
468
User Manual 1326.1578.02 ─ 12
Parameters:
<ChecksumError>
ON | OFF
*RST:
ON
TRIGger:LIN:DATA
<Pattern>
Defines the data pattern as trigger condition. Enter the words in MSB first bit order.
Parameters:
<Pattern>
String with max. 4 bytes in binary format. Characters 0, 1, and X
are allowed. If you define a pattern with incomplete byte, the
missing LSB are filled with X.
Example:
TRIG:LIN:TYPE IDDT
TRIG:LIN:IDEN 001001
TRIG:LIN:ICON EQU
TRIG:LIN:DCON EQU
TRIG:LIN:DATA '11111111000000001111'
TRIG:LIN:DATA?
<-- 11111111000000001111XXXX
Triggers when the id '001001' and the data
'11111111000000001111' is detected in the measured signal
TRIGger:LIN:DCONdition
<Relation>
Defines how the specified data pattern is compared with the acquired signal.
Parameters:
<Relation>
EQUal | NEQual
*RST:
EQUal
TRIGger:LIN:ICONdition
<Relation>
Defines how the specified identifier pattern is compared with the acquired signal. The
instrument triggers if the acquired address is equal or unequal the defined pattern.
Parameters:
<Relation>
EQUal | NEQual
*RST:
EQUal
TRIGger:LIN:IDENtifier
<Pattern>
Specifies the identifier pattern to be found, in binary format. Enter the pattern in MSB
first bit order.
Parameters:
<Pattern>
String with max. 7 characters. Characters 0, 1, and X are
allowed. If you define a pattern shorter than the pattern length,
the missing LSB are filled with X.