Remote Control Commands
R&S
®
Scope Rider RTH
301
User Manual 1326.1578.02 ─ 12
Suffix:
<m>
.
1|2 (RTH1002) , 1..4 (RTH1004)
Parameters:
<ProbeSettings>
V1TO1 | V10To1 | V20To1 | V100to1 | V200to1 | V1000to1 |
C100V1A | C10V1A | C1V1A | C100MV1A | C10MV1A |
C1MV1A | T1MVC | T1MVF | PT100 | PT500 | PT1000 | USER
V1TO1 | V10To1 | V20To1 | V100To1 | V200To1 | V1000To1
Attenuation factors of voltage probes: 1:1, 10:1, 20:1, 100:1,
200:1, 1000:1.
C100V1A | C10V1A | C1V1A | C100MV1A | C10MV1A |
C1MV1A
Sensitivity of current probes: 100 V/A, 10 V/A, 1 V/A, 100 mV/A,
10 mV/A, 1 mV/A.
T1MVC | T1MVF | TPTC | TPTF
Temperature adapter type: 1 mV/°C, 1 mV/°F, PT100(°C),
PT100(°F)
USER
Set a user-defined attenuation factor if the predefined values do
not fit: Use
PROBe<m>:SETup:ATTenuation:MANual
PROBe<m>:SETup:ATTenuation:UNIT
to set
the unit.
*RST:
V1TO1
PROBe<m>:SETup:ATTenuation:MANual
<ProbeFactor>
Sets a user-defined attenuation factor if
is set to
USER
.
Suffix:
<m>
.
1|2 (RTH1002) , 1..4 (RTH1004)
Parameters:
<ProbeFactor>
Range:
100E-6 to 10000
Increment: 100E-6
*RST:
1
PROBe<m>:SETup:ATTenuation:UNIT
<ProbeUnit>
Sets a user-defined probe unit if
USER
.
Suffix:
<m>
.
1|2 (RTH1002) , 1..4 (RTH1004)
Parameters:
<ProbeUnit>
V | A
*RST:
V
Waveform Setup