Remote Control Commands
R&S
®
Scope Rider RTH
455
User Manual 1326.1578.02 ─ 12
TRIGger:CAN:CRCerror
<CrcError>
A CRC error occurs when the calculated result differs from the received value in the
CRC sequence.
Parameters:
<CrcError>
ON | OFF
*RST:
ON
TRIGger:CAN:DATA
<Pattern>
Specifies the data pattern to be found, in binary or hex format. Enter the pattern in
MSB first bit order.
Parameters:
<Pattern>
String that contains the pattern in binary format. The parameter
accepts the bit value X (don't care).
TRIGger:CAN:DCONdition
<Relation>
Defines how the specified data pattern is compared with the acquired signal. The
instrument triggers if the acquired address is equal or unequal the defined pattern.
Parameters:
<Relation>
EQUal | NEQual
*RST:
EQUal
TRIGger:CAN:FORMerror
<FormError>
A form error occurs when a fixed-form bit field contains one or more illegal bits.
Parameters:
<FormError>
ON | OFF
*RST:
ON
TRIGger:CAN:FTYPe
<FrameType>
CAN has several frame types which can be used as trigger condition.
For data and remote frames, the identifier format has to be set with
.
Parameters:
<FrameType>
ERRor | OVERload | DATA | REMote | DOR
ERRor
When a node recognizes an error, it cancels transmission by
sending an error frame.
The instrument triggers seven bit periods after the end of the
error flag that is marked by a dominant-recessive edge.
Protocol Analysis