Remote Control Commands
R&S
®
Scope Rider RTH
378
User Manual 1326.1578.02 ─ 12
HARMonic:RESult<m>:THD:MAXimum?
HARMonic:RESult<m>:THD:MINimum?
HARMonic:RESult<m>:THD[:CURRent]?
Returns the currently calculated, minimum, or maximum THD (total harmonic distor-
tion), that is: the RMS of the amplitude of the harmonics. Which value the THD is set in
relation to depends on the
Suffix:
<m>
.
1|2 (RTH1002) , 1..4 (RTH1004)
Return values:
<THD>
Range:
0 to 1000
Increment: 0.1
*RST:
0
Default unit: %
Example:
HARM:THDT THDF
HARM:RES1:THD?
Returns the RMS amplitude (voltage or current) of the harmon-
ics relative to the RMS amplitude of the fundamental compo-
nent.
Usage:
Query only
15.7
Digital Multimeter (R&S
RTH1002)
Activating the Multimeter mode
.............................................................................378
..................................................................................379
........................................................................................ 386
............................................................................................388
15.7.1
Activating the Multimeter mode
To activate the DMM, use
OP METer
.
OP[:MODE]
<OperationMode>
Sets the operating mode of the instrument.
Parameters:
<OperationMode>
YT | XY | MASK | ROLL | METer | FFT | LOGGer | COUNter |
PROTocol | HARMonic | SPECtrum
*RST:
YT (scope mode)
Digital Multimeter (R&S
RTH1002)