Remote Control Commands
R&S
®
Scope Rider RTH
400
User Manual 1326.1578.02 ─ 12
METer<m>:SENSe:VOLTage:AC:NULL:VALue
..................................................................400
METer<m>:SENSe:VOLTage:DC:NULL:VALue
..................................................................400
.......................................................................................401
METer<m>:SENSe:CURRent:AC:NULL:STATe
<State>
METer<m>:SENSe:CURRent:DC:NULL:STATe
<State>
METer<m>:SENSe:VOLTage:AC:NULL:STATe
<State>
METer<m>:SENSe:VOLTage:DC:NULL:STATe
<State>
Enables or disables the relative measurement. The reference value is defined using
METer:SENSe:<function>:NULL:VALue
Suffix:
<m>
.
1..4
Selects the voltmeter.
Parameters:
<State>
OFF | ON
*RST:
OFF
METer<m>:SENSe:NULL:STATe
<SetRelative>
METer<m>:SENSe:RELative:STATe
<SetRelative>
Enables or disables the relative measurement for the currently active measurement
type.
Relative measurements are not available for diode, continuity, and frequency measure-
ments.
Suffix:
<m>
.
1..4
R&S
RTH1002: always 1, omit the suffix
R&S
RTH1004: Selects the voltmeter.
Parameters:
<SetRelative>
ON | OFF
ON = 1, OFF = 0
*RST:
OFF
METer<m>:SENSe:CURRent:AC:NULL:VALue
<ReferenceValue>
METer<m>:SENSe:CURRent:DC:NULL:VALue
<ReferenceValue>
METer<m>:SENSe:VOLTage:AC:NULL:VALue
<ReferenceValue>
METer<m>:SENSe:VOLTage:DC:NULL:VALue
<ReferenceValue>
Sets the reference value for relative measurements. The measurement result is the dif-
ference of the measured sample and the reference value.
The value takes effect if relative measurement is on, see
.
Voltmeter (R&S
RTH1004)