General Instrument Setup
R&S
®
Scope Rider RTH
282
User Manual 1326.1578.02 ─ 12
The R&S
RTH1002 provides an additional, separate self-alignment of the meter inputs.
Preparing the instrument for self-alignment
Make sure that the instrument has been running and warming up before you start the
self-alignment. The minimum warm-up time is indicated in the data sheet.
Before the self-alignment, remove all probes, leads, and other connected lines from the
instrument input.
Figure 13-1: Self-alignment in scope mode
► To start, tap "Start Selfalignment".
The alignment can take up to 15 minutes. The results are shown in the "Selfalign-
ment" dialog box.
R&S
RTH1002: Self-alignment of DMM inputs
If the instrument is in "Meter" mode, you can choose to align all inputs, or only the
meter inputs. The self-alignment of the meter inputs takes up to 30 seconds, while the
complete self-alignment takes up to 15 minutes.
Selfalignment