Remote Control Commands
R&S
®
Scope Rider RTH
488
User Manual 1326.1578.02 ─ 12
LOGic:GROup<m>:THReshold?
Returns the current threshold value.
Suffix:
<m>
.
1..3
1 = all logic channels D0 to D7
2 = group D0 to D3
3= group D4 to D7
Return values:
<Level>
Range:
-10 to 10
Increment: 1E-3
*RST:
0
Default unit: V
Usage:
Query only
LOGic:GROup<m>:HYSTeresis
<Hysteresis>
Hysteresis avoids the change of signal states due to noise oscillation around the
threshold level. Set a small hysteresis for clean signals, and large hysteresis for noisy
signals.
Suffix:
<m>
.
1..3
1 = all logic channels D0 to D7
2 = group D0 to D3
3= group D4 to D7
The suffix only takes effect if
OFF
.
Parameters:
<Hysteresis>
SMALl | MEDium | LARGe
*RST:
MEDium
LOGic:CHANnel<m>:DESKew
<Value>
Sets the deskew for all channels of a logic probe at once, or for each logic channel
separately.
Suffix:
<m>
.
1..9
1..8: logic channels 0 to 7
9: all logic channels
Parameters:
<Value>
Range:
-100E-9 to 100E-9
Increment: 800E-12
*RST:
0
Default unit: s
Logic Analyzer (R&S
RTH-B1 MSO)