Protocol Analysis
R&S
®
Scope Rider RTH
234
User Manual 1326.1578.02 ─ 12
9.7.3
SENT Trigger Settings
Access: [Setup] ([Trigger]) > "Trigger type" = "Bus"
SENT Trigger
Sets the SENT trigger type. Depending on the selected value, different additional
parameters are available.
"Start of Frame"
Triggers at the end of the synchronization/calibration pulse.
"Fast status"
Sets the trigger to a specific status or a specific pattern from the fast
channel.
"Fast status and data"
Sets the trigger to a combination of status and data condition from the
fast channel. Triggers at the end of the last data nibble.
"Slow Identifier"
Sets the trigger to a specific identifier or a specific pattern from the
slow channel. The instrument triggers at the end of the last transmis-
sion sequence which includes the last serial bits of the slow channel.
SENT (Option R&S
RTH-K10)