Waveform Setup
R&S
®
Scope Rider RTH
68
User Manual 1326.1578.02 ─ 12
Remote command:
Slope
Sets the edge direction for the trigger. You can trigger on:
●
rising edge, that is a positive voltage change
●
falling edge, that is a negative voltage change
●
rising and falling edge
Remote command:
Range
Defines the time limits of the slew rate. The comparisons are the same as for the width
trigger, see
Remote command:
Time
Sets the slew rate for comparisons equal, unequal, shorter, and longer.
Remote command:
±Tolerance
Sets a tolerance range
Δ
t to the specified
if the comparison range is equal or
unequal.
Remote command:
Min Time / Max Time
Set the lower and upper time limits if "Inside" or "Outside" is set for comparison.
Remote command:
3.6.11
Data2Clock Trigger (R&S
RTH-K19)
With the Data2Clock trigger - also known as setup/hold trigger - you can analyze the
relative timing between two signals: a data signal and the synchronous clock signal.
Many systems require, that the data signal must be steady for some time before and
after the clock edge, for example, the data transmission on parallel interfaces.
The reference point for the time measurement is defined by clock level and clock edge.
► [SETUP] key > "Trigger Type" = "Data2Clk"
Trigger