Remote Control Commands
R&S
®
Scope Rider RTH
320
User Manual 1326.1578.02 ─ 12
LONGer | SHORter
Triggers on pulses shorter or longer than a runt width that is
defined by
.
EQUal | NEQual
Triggers pulses with a width equal or unequal to a given width
and an optional tolerance defined by
WITHin | OUTSide
Triggers on pulses inside or outside a given range. The range is
defined by ..
*RST:
LONGer
TRIGger:PATTern:TIMeout[:TIME]
<Time>
Sets a minimum time during which the signals match the pattern definition defined by
TRIGger:PATTern:STATe[:CHANnel<m>]
and
.
The command is required if
is set to TIMeout.
Parameters:
<Time>
Range:
100E-12 to 10000
Increment: 100E-9
*RST:
5E-9
Default unit: s
TRIGger:PATTern:WIDTh[:WIDTh]
<Duration>
Sets the width for comparison ranges LONGer, SHORter, EQUal and NEQual.
See
Parameters:
<Duration>
Range:
800E-12 to 10000
Increment: 100E-9
*RST:
5E-9
Default unit: s
TRIGger:PATTern:WIDTh:DELTa
<Tolerance>
Sets a range
Δ
t to the pattern width that is defined by
.
Parameters:
<Tolerance>
Range:
0 to 5000
Increment: 500E-12
*RST:
0
Default unit: s
Waveform Setup