Remote Control Commands
R&S
®
Scope Rider RTH
369
User Manual 1326.1578.02 ─ 12
HARMonic:FUNDamental:USER
<UserFrequency>
Defines the user-defined fundamental frequency if
is
set to
USER
.
Parameters:
<UserFrequency>
Range:
10 to 1E+6
Increment: 0.1
*RST:
50
Default unit: Hz
Example:
HARMonic:FUNDamental:TYPE USER
HARMonic:FUNDamental:USER 123
HARMonic:FUNDamental:CURRent?
//Result: 123
HARMonic:FUNDamental:CURRent?
Returns the current fundamental frequency.
Return values:
<CurrFreq>
Range:
10 to 1E+6
Increment: 0.01
*RST:
50
Default unit: Hz
Usage:
Query only
HARMonic:LIMits:NAME
<FileName>
Parameters:
<FileName>
String containing the path and name of the user-defined configu-
ration file to be loaded.
Example:
HARM:LIM:NAME
'/media/SD/Rohde-Schwarz/RTH/Harmonic/LimitExample.csv'
HARM:LIM:LOAD
HARMonic:LIMits:LOAD
Loads the file selected by
Example:
HARM:LIM:NAME
'/media/SD/Rohde-Schwarz/RTH/Harmonic/LimitExample.csv'
HARM:LIM:LOAD
Usage:
Event