For Your Safety
R&S
®
Scope Rider RTH
8
User Manual 1326.1578.02 ─ 12
Battery replacement
●
Disconnect power supply, probes, test leads and all other cables before opening
the battery cover.
●
Use only the specified Li-ion battery pack, which is delivered with the instrument.
You can order additional battery packs at Rohde
&
Schwarz, see data sheet for
order number.
●
Do not operate the instrument with the battery cover open.
●
Use only the specified power adapter, which is delivered with the instrument.
Cleaning
Remove all probes, leads, USB and LAN cables and power supply before cleaning the
instrument. Use only cleaning materials that are specified in the manual.
Measurement categories
IEC 61010-2-030 defines measurement categories that rate instruments on their ability
to resist short transient overvoltages that occur in addition to the working voltage. Use
the measurement instrument and accessories only in electrical environments for which
they are rated.
●
O - Instruments without rated measurement category
For measurements performed on circuits not directly connected to mains, for exam-
ple, electronics, circuits powered by batteries, and specially protected secondary
circuits. This measurement category is also known as CAT I.
●
CAT II:
For measurements performed on circuits directly connected to the low-voltage
installation by a standard socket outlet, for example, household appliances and
portable tools.
●
CAT III:
For measurements performed in the building installation, such as junction boxes,
circuit breakers, distribution boards, and equipment with permanent connection to
the fixed installation.
●
CAT IV:
For measurements performed at the source of the low-voltage installation, such as
electricity meters and primary overcurrent protection devices.