Remote Control Commands
R&S
®
Scope Rider RTH
384
User Manual 1326.1578.02 ─ 12
Parameters:
<Range>
<numeric value> | MIN | MAX | DEF
<numeric value>
The instrument has fixed measurement ranges: 1 V|10 V|100 V|
1000
V.
You can enter any value between the minimum and maximum
value. The instrument uses the next suitable range.
Range:
1 to 1000
*RST:
1
Default unit: V
15.7.2.3
METer:SENSe:<function>:RANGe:AUTO Commands
METer:SENSe:<function>:RANGe:AUTO
commands turn autoranging on or off. For
some measurements, you can also use the
METer:CONFigure:<function>
com-
Chapter 15.7.2.1, "METer:CONFigure Commands"
METer:SENSe:CAPacitance:RANGe:AUTO
......................................................................384
METer:SENSe:CURRent:AC:RANGe:AUTO
..................................................................... 384
METer:SENSe:CURRent:DC:RANGe:AUTO
..................................................................... 384
METer:SENSe:FREQuency:VOLTage:RANGe:AUTO
.........................................................384
METer:SENSe:RESistance:RANGe:AUTO
........................................................................384
METer:SENSe:VOLTage:AC:RANGe:AUTO
......................................................................384
METer:SENSe:VOLTage:DC:RANGe:AUTO
......................................................................384
METer:SENSe:CAPacitance:RANGe:AUTO
<State>
METer:SENSe:CURRent:AC:RANGe:AUTO
<State>
METer:SENSe:CURRent:DC:RANGe:AUTO
<State>
METer:SENSe:FREQuency:VOLTage:RANGe:AUTO
<State>
METer:SENSe:RESistance:RANGe:AUTO
<State>
METer:SENSe:VOLTage:AC:RANGe:AUTO
<State>
METer:SENSe:VOLTage:DC:RANGe:AUTO
<State>
Disables or enables the autoranging for the specified measurement.
The query always returns
OFF
or
ON
.
Parameters:
<State>
OFF | ON | ONCE
OFF | ON
0 | 1 are not supported.
ONCE
Performs an immediate autorange and then turns off the auto-
ranging.
Example:
METer:SENSe:VOLTage:DC:RANGe:AUTO ONCE
METer:SENSe:VOLTage:DC:RANGe:AUTO?
<-- OFF
Digital Multimeter (R&S
RTH1002)