Protocol Analysis
R&S
®
Scope Rider RTH
222
User Manual 1326.1578.02 ─ 12
LIN Trigger
Depending on the selected LIN trigger type, different additional parameters are availa-
ble.
"Start of Frame"
Triggers on the stop bit of the sync field.
"Wakeup frame"
Triggers after a wakeup frame.
"Error condition"
Identifies various errors in the frame. You can select one or more
error types as the trigger condition.
"Identifier"
Sets the trigger to a specific identifier or an identifier range. Only the
6-bit identifier without parity bits is considered, not the protected iden-
tifier.
"Iden Data"
Sets the trigger to a combination of identifier and data condition. The
instrument triggers at the end of the last byte of the specified data
pattern.
Remote command:
Checksum Error
Triggers on a checksum error. The checksum verifies the correct data transmission. It
is the last byte of the frame response. The checksum includes not only the data but
also the protected identifier (PID).
Remote command:
Parity Error
Triggers on a parity error. Parity bits are the bits 6 and 7 of the identifier. They verify
the correct transmission of the identifier.
Remote command:
Sync Error
Triggers if synchronization caused an error.
Remote command:
LIN (Option R&S