Selects a user-defined number of harmonics. Define the number
of harmonics using the
command.
*RST:
ALL
HARMonic:DISPlay:USER
<MaxHarmonics>
Determines the maximum number of harmonics to be displayed in the bargraph. This
command requires a preceding
command.
Note that this setting only affects the graphical result display; it has no effect on the
selection of harmonics for which results are provided during a file export (manually or
via remote command). For that purpose, use
Harmonics that are eliminated by the
displayed.
Parameters:
<MaxHarmonics>
Range:
1 to 64
Increment: 1
*RST:
10
Example:
HARM:DISP:TYPE USER
HARM:DISP:USER 25
Displays the first 25 harmonics.
HARMonic:FUNDamental:TYPE
<Fundamental>
Defines the basis of the harmonics measurement. Harmonics are determined as multi-
ples of this frequency.
Parameters:
<Fundamental>
F50 | F60 | F400 | USER
F50
50 Hz
F60
60 Hz
F400
400 Hz
USER
User-defined frequency; define the frequency using
*RST:
F50
Example:
HARMonic:FUNDamental:TYPE USER
HARMonic:FUNDamental:USER 123
HARMonic:FUNDamental:CURRent?
//Result: 123
Spectrum Analysis