Multimeter Measurements
R&S
®
Scope Rider RTH
146
User Manual 1326.1578.02 ─ 12
Figure 7-2: Display of the digital multimeter
1 = Measurement state. "Manual" or "Auto": running measurement with manual or automatic range. "Hold":
stopped measurement.
2 = Measurement range
3 = Test lead connection including polarity
4 = Minimum value and timestamp
5 = Average value
6 = Maximum value and timestamp
7 = Warning sign, shows up if the measured value is higher than 30 V
8 = Bargraph displaying the measured value
9 = Measurement type
10 = Activates or deactivates relative measurement
11 = Restarts the measurement and resets all values
In meter mode, the keys behave slightly different from scope mode:
●
The AUTOSET key toggles the range mode: autoranging and manual range set-
ting.
●
The vertical [RANGE] and [POS] keys adjust the measurement range.
●
The [MEAS] key opens the "Meter" menu.
●
The following keys work as usual: [FILE],
,
, [PRESET], [MODE], [BACK].
●
The [CH] keys switch back to the scope mode.
●
All other keys do not work.
7.1.4
Running Multimeter Measurements
1. Connect the test leads to the meter inputs.
2. Select the "Meter" mode.
3. Select the measurement type (no. 9 in
Digital Multimeter (R&S
RTH1002)