Command reference
R&S
®
ZNL/ZNLE
951
User Manual 1178.5966.02 ─ 19
Bandwidths can be set in 1
–
1.5
–
2
–
3
–
5
–
7 steps. The analyzer rounds up any
entered value between these steps and rounds down values exceeding the maximum
bandwidth.
At the same time, the command activates separate bandwidth setting in all sweep seg-
ments (
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]:CONTrol
ON
).
Suffix:
<Ch>
.
Channel number
<Seg>
Sweep segment number
Parameters:
<ResBandwidth>
IF bandwidth
Range:
See above
Increment: 1-1.5-2-3-5-7 steps
*RST:
10 kHz
Default unit: Hz
Example:
See
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:
Manual operation:
See
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]:CONTrol
<Boolean>
Selects common or independent "Meas. Bandwidth" settings for the sweep segments.
Suffix:
<Ch>
.
Channel number
<Seg>
Sweep segment number. This suffix is ignored; the setting con-
trols the whole segmented sweep.
Parameters:
<Boolean>
ON
–
use independent bandwidth settings, to be defined via
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]
OFF
–
reset the bandwidth in all sweep segments to the band-
width for unsegmented sweeps, defined via
. ON will not restore the previous val-
ues.
The parameter is automatically switched to ON when a band-
width is entered using
*RST:
OFF
VNA command reference