VNA GUI reference
R&S
®
ZNL/ZNLE
341
User Manual 1178.5966.02 ─ 19
Delay
Specifies a delay time between the trigger event and the start of the next measurement
sequence.
The specified "Delay" must be zero or positive, so that the trigger event precedes the
start of the measurement (post-trigger).
If "Multiple Triggers" is active, the "Delay" can be selected individually using the
Remote command:
TRIGger<Ch>[:SEQuence]:HOLDoff
Signal Type
Specifies the "External" trigger mode in detail.
●
"Rising Edge"/"Falling Edge"
means that the rising/falling slope of every external
trigger pulse can trigger a single measurement sequence.
●
"Active High"/"Active Low"
means that the analyzer measures in "FreeRun"
mode as long as the external trigger signal is high/low. The measurement is dis-
continued when the trigger signal changes to low/high.
Remote command:
10.5.3.2
Trigger Manager dialog
The "Trigger Manager" dialog defines individual trigger sources and delays for the trig-
gered measurement sequences. The dialog is available and its settings are valid if the
analyzer is configured for "Multiple Triggers" (see
"FreeRun / External / Manual / Multi-
Access:
[Sweep] > "Trigger" > "Trigger Manager..."
Background information
Refer to
Chapter 9.1.4.1, "Partial measurements and driving mode"
Sweep Softtool