Command reference
R&S
®
ZNL/ZNLE
937
User Manual 1178.5966.02 ─ 19
Example:
*RST; CORR:POW:ACQ AWAVe,1,PORT,1
Perform a receiver power calibration of the wave a
1
using port 1
as a source port, assuming that the transmitted source power a
1
is correct. No external cabling is needed.
CORR:POW:AWAV?
Check whether the calibration is applied (the response is
1
).
CORR:PCAL NONE; POW:AWAV?
Disable all power calibrations and check again whether the cali-
bration is applied (the response is
0
).
Usage:
Setting only
Manual operation:
See
"Scalar Power Cal – All Power Cals On / All Power Cals
[SENSe<Ch>:]CORRection:POWer<PhyPt>:ACQuire
<Wave>[, <CalPort>[,
<SourceType>[, <SourcePort>[, <CalOnlyPortFreq>]]]]
Selects the wave quantity and the source for the receiver power calibration, starts the
calibration sweep, and applies the receiver power correction.
Suffix:
<Ch>
.
Calibrated channel number
<PhyPt>
This suffix is ignored because the port number is specified in the
parameter list.
Setting parameters:
<Wave>
AWAVe | BWAVe | B1 | B2 | B3 | B4
AWAVe
–
calibration of reference waves a
1
, a
2
, .... In manual
control, the reference receiver calibration is included in the
source power calibration of each port.
BWAVe
–
calibration of received wave b
1
, b
2
, .... The port num-
ber
<CalPort>
, the used
<SourceType>
, and the
<SourcePort>
number must be specified in addition.
Alternative: Parameters B1 and B2.
B1 | B2 ...
–
direct wave and cal port setting for received waves
b
1
, b
2
... The parameters B1 and B2 are ZVR-compatible. No
additional parameters must be specified. The source for B1 is
port 2 and vice versa.
<CalPort>
Calibrated port number
Range:
1 to port number of the analyzer
<SourceType>
PORT
PORT - internal source at port
<CalPort>
, to be fed to port
<SourcePort>
using an external through connection. If
<CalPort> = <SourcePort>
, an Open or Short standard is
required.
VNA command reference