VNA GUI reference
R&S
®
ZNL/ZNLE
388
User Manual 1178.5966.02 ─ 19
On loading data from a trace file with several traces, the analyzer displays a dialog to
select one or more of the traces stored in the file (see
on page 393). E.g., for an
*.s2p
Touchstone file, the box offers all
four 2-port S-parameters (see
Chapter 9.4.2.1, "Touchstone files"
Coupling between the imported memory trace and the active data trace implies that the
stimulus values of the imported data and of the active trace must be compatible. Com-
patibility means that the "Sweep Type" of the two traces must match; the position and
number of the sweep points do not have to be the same.
The analyzer checks for compatibility before importing data. The "Select Parameter"
box remains empty if the selected file contains no compatible data.
"Import Complex Data" is a standard "Open File" dialog with an additional button.
Import Data to New Mem
Specifies whether the loaded data overwrite an existing memory trace, if available (box
unchecked), or whether they are used to generate a new memory trace (box checked).
If the box is unchecked and the active trace is a memory trace, then this memory trace
will be overwritten. If the box is unchecked and the active trace is a data trace, then the
data trace's last created memory trace will be overwritten (or a new memory trace will
be created, in case there was previously no memory trace assigned to this data trace).
Remote command:
10.6.11.2
Export Data - <File Type> dialog
The "Export Data - <File Type>" dialog stores data or memory traces to a trace file.
Trace files are ASCII files with selectable file format.
Access:
●
[Trace] > "Trace Data" > "Export snp Files" – ...
●
[Trace] > "Trace Data" > "Export" – ...
Trace softtool