Command reference
R&S
®
ZNL/ZNLE
827
User Manual 1178.5966.02 ─ 19
<WndTr>
Existing trace number, assigned by means of
WINDow<Wnd>]:TRACe<WndTr>:FEED
. This suffix is ignored if
the optional
<TraceName>
parameter is used.
Setting parameters:
<Activate>
ONCE
Activate the autoscale function.
<TraceName>
Optional string parameter for the trace name, e.g. 'Trc4'. If this
optional parameter is present, both numeric suffixes are ignored
(trace names must be unique across different channels and win-
dows).
Example:
*RST; DISP:WIND:TRAC:Y:PDIV?; RLEV?
Query the value between two grid lines and the reference value
for the default trace. The response is
10;0
.
DISP:WIND:TRAC:Y:AUTO ONCE; PDIV?; RLEV?
or:
DISP:WIND:TRAC:Y:AUTO ONCE, 'Trc1'; PDIV?;
RLEV?
Autoscale the default trace and query the scaling parameters
again. In general both values have changed.
Usage:
Setting only
Manual operation:
See
DISPlay[:WINDow<Wnd>]:TRACe<WndTr>:Y[:SCALe]:BOTTom
<LowEdge>[,
<TraceName>]
DISPlay[:WINDow<Wnd>]:TRACe<WndTr>:Y[:SCALe]:TOP
<UppEdge>[,
<TraceName>]
These commands define the lower (bottom) and upper (top) edge of the diagram area
<Wnd>
.
Suffix:
<Wnd>
.
Number of an existing diagram area (defined by means of
ON
). This suffix is
ignored if the optional
<TraceName>
parameter is used.
<WndTr>
Existing trace number, assigned by means of
WINDow<Wnd>]:TRACe<WndTr>:FEED
. This suffix is ignored if
the optional
<TraceName>
parameter is used.
Parameters:
<UppEdge>
Value and unit for the lower or upper diagram edge. Range and
unit depend on the measured quantity, see
Default unit: NN
VNA command reference