VNA GUI reference
R&S
®
ZNL/ZNLE
545
User Manual 1178.5966.02 ─ 19
If "Delay" is the selected trace format, the entire trace is shifted in vertical direction and
centered on zero. In phase format, the "Auto Length" corrected trace shows the devia-
tion from linear phase.
Use
to limit the frequency range for the calculation.
Note
that in order to calculate the electrical length offset of port
i
, you have to select a
trace that uses port
i
as one of its receive ports. It is
not
relevant for the "Auto Length"
If the measured quantity is a ratio, or if it is derived from a ratio, its receiving port is
given as the index of the wave quantity in the numerator. If the active trace shows an
S-parameter S
ij
, then "Auto Length" adds a length offset at port i. If the receive port is
balanced, then the same offset is applied to both its physical ports.
Remote command:
[SENSe<Ch>:]CORRection:EDELay:AUTO
Fixture Compensation...
The "Fixture Compensation..." button opens the
.
Evaluation Range...
The "Evaluation Range..." button opens the
that allows you to
restrict the frequency range of auto length (and loss) calculations.
10.14.2.2
Fixture Compensation dialog
This dialog allows you to correct the measurement result for the effects of a test fixture.
Access:
●
[Offset
Embed] > "Offset" > "Fixture Compensation..."
●
[Offset
Embed] > "One Way Loss" > "Fixture Compensation..."
Background Information
Refer to
Chapter 9.6.1.5, "Fixture Compensation"
Offset Embed softtool