Command reference
R&S
®
ZNL/ZNLE
721
User Manual 1178.5966.02 ─ 19
CALCulate<Chn>:RIPPle:DISPlay[:STATe]
<Boolean>
Displays or hides all ripple limit lines (including all ranges) associated to the active
trace.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean>
ON | OFF - ripple limit line on or off.
*RST:
OFF
Example:
*RST; CALC:RIPP:DATA 1,1500000000, 2000000000,
3
Define and enable a ripple limit range in the stimulus range
between 1.5 GHz and 2 GHz, assigning a ripple limit of +3 dB.
CALC:RIPP:DISP ON
Show the ripple limit range in the active diagram.
Manual operation:
See
CALCulate<Chn>:RIPPle:FAIL?
Returns a 0 or 1 to indicate whether or not the global ripple limit check has failed.
Tip:
CALCulate<Chn>:RIPPle:SEGMent<Seg>:RESult?
to query the result
for a single ripple limit range.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Example:
*RST; CALC:RIPP:DATA 1, 1500000000, 2000000000,
3
Define and enable a ripple limit range in the stimulus range
between 1.5 GHz and 2 GHz, assigning a ripple limit of +3 dB.
CALC:RIPP:STAT ON; FAIL?
Switch the limit check on and query the result.
CALC:RIPP:STAT:AREA LEFT, TOP
For a subsequent check at the GUI or a hardcopy, move the info
field to the top left position.
Usage:
Query only
Manual operation:
See
CALCulate:RIPPle:FAIL:ALL?
[<ChannelSetup>]
Returns a 0 or 1 to indicate whether or not the global ripple limit check has failed for at
least one channel in the referenced channel setup.
Query parameters:
<ChannelSetup>
Name of a VNA channel setup. Can be omitted if the active VNA
channel setup shall be referenced.
VNA command reference