VNA GUI reference
R&S
®
ZNL/ZNLE
483
User Manual 1178.5966.02 ─ 19
●
The full n-port calibration types (n > 1, e.g. "TOSM") are symbolized by a closed
square symbol. The number of arrows increases the complexity but can also
improve the accuracy of the calibration.
For an overview, refer to
Remote command:
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine
Calibrate all Channels
Check this box to apply the acquired correction data to all channels in the active chan-
nel setup. Leave it unchecked (preset setting) to apply them only to the active channel.
Note that this option is available only if the active channel setup contains multiple
channels.
Remote command:
[SENSe:]CORRection:COLLect:CHANnels:ALL
Next
Step 2: Connectors and Cal Kits
Step 2: Connectors and Cal Kits
Selects the connector type and gender for all ports and allows you to import a calibra-
tion kit.
Background information
Refer to
Chapter 9.5.2, "Calibration standards and calibration kits"
Messages in the dialog
An information message (or error message) is displayed if one of the following hap-
pens:
●
One of the selected calibration kits is described by ideal kit parameters or typical
values.
●
One of the selected calibration kits does not contain all standards that are required
for the previously selected calibration type.
●
Different connector types are defined at the ports but the selected calibration type
requires uniform connectors.
●
A cal kit standard does not cover the entire calibrated frequency range.
Cal softtool