Command reference
R&S
®
ZNL/ZNLE
810
User Manual 1178.5966.02 ─ 19
Query parameters:
<TraceName>
Trace name (string), e.g.
'Trc1'
Example:
*RST; :CALC:PAR:SDEF 'Trc2', 'S11'
Create a trace named Trc2.
CONF:TRAC:WIND:TRAC? 'Trc2'
Query the diagram number for Trc2. The new trace is not dis-
played, so the response is
0
.
DISP:WIND2:STAT ON
Create a diagram no. 2.
DISP:WIND2:TRAC3:FEED 'Trc2'
Display the trace in the new diagram no. 2, assigning the trace
number 3.
CONF:TRAC:WIND? 'Trc2'
Query the diagram number for Trc2. The response is
2
.
CONF:TRAC:WIND:TRAC? 'Trc2'
Query the trace number for Trc2. The response is
3
.
Usage:
Query only
Manual operation:
See
11.5.2.3
CONTrol commands
The
Control...
commands allow you to configure the following interfaces:
●
Aux. Port
●
The VNA User Port functionality is offered at the Aux. Port connector that is provi-
ded by hardware option R&S
FPL1-B5 "Additional Interfaces".
This option is not available for the R&S
ZNLE
●
Usage of the Aux. Port as VNA User Port (instead of SA Auxiliary Port) must be
selected in the "Add. Interfaces" tab of the "System Configuration" dialog.
........................................................................................... 810
CONTrol:AUXiliary:C[:DATA]
<DecValue>
Sets or queries a channel-dependent eight-bit decimal value to control eight independ-
ent output signals at the Aux. Port connector (lines 8, 9, 10, 11 and lines 16, 17, 18,
19). The output signals are 3.3 V TTL signals which can be used to differentiate
between up to 255 independent analyzer states.
CONTrol:AUXiliary:C[:DATA]
itself does not change the analyzer state.
Channel bit definition and activation
The channel bits have the following properties:
●
After a
*RST
of the analyzer all channel bits (including the value for the active,
sweeping channel no. 1) are set to zero; no signal is applied to pins 8 to 11 and 16
to 19 of the Aux. Port connector.
VNA command reference