Command reference
R&S
®
ZNL/ZNLE
665
User Manual 1178.5966.02 ─ 19
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<CenterX>
Range:
Virtually no restriction for center coordinates.
*RST:
0
Default unit: NN
<CenterY>
Range:
Virtually no restriction for center coordinates.
*RST:
0
Default unit: NN
<Radius>
Range:
Virtually no restriction for radius (use positive val-
ues).
*RST:
1
Default unit: NN
Example:
See
CALCulate<Chn>:LIMit:CIRCle[:STATe]
Manual operation:
See
"Radius / Center X / Center Y"
CALCulate<Chn>:LIMit:CIRCle:DISPlay[:STATe]
<Boolean>
Displays or hides the circle limit line associated to the active trace.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean>
ON | OFF - Circle limit line on or off.
*RST:
OFF
Example:
*RST; CALCulate:LIMit:CIRCle:DATA 0, 0, 0.5
Define a circle limit line centered around the origin of the polar
diagram, assigning a radius of 0.5 U.
CALCulate:FORMat POLar
CALCulate:LIMit:CIRCle:DISPlay ON
Activate a polar diagram and show the circle limit line in the dia-
gram.
Manual operation:
See
CALCulate<Chn>:LIMit:CIRCle:FAIL?
Returns a 0 or 1 to indicate whether or not the circle limit check has failed. 0 repre-
sents pass and 1 represents fail
Tip:
to perform a composite (global) limit check.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Example:
See
CALCulate<Chn>:LIMit:CIRCle[:STATe]
Usage:
Query only
VNA command reference