414
Appendix A
Command Reference
LMIN
Digital Video Minimum quantizing Level
Class
Format parameter setting - Digital video signal
Description
Sets the minimum quantizing level of the component video signal for black. The default is
0 for black; however, some applications may require that the minimum level be set higher.
The query reads the current setting for the minimum level.
Command syntax
LMIN
video_level
video_level
unsigned integer from 0 to 4 when NBPC is 6 gamuts; 0 to 16 when NBPC is 8
gamuts; 0 to 64 when NBPC is 10 gamuts; 0 to 256 when NBPC is 12 gamuts
Query syntax
LMIN?
Example
LMIN 16 // To set the minimum digital quantizing level to 16
Related commands
LMAX
,
NBPC
,
DVQM
Summary of Contents for 802BT