440
Appendix A
Command Reference
MSIZ
light Meter SIZe
Class
System parameter settings
Description
Establishes the physical size of the light meter boxes displayed in the BriteBox test image.
The unit of measure used is based on the current setting of the system level
USIZ
command. It also affects the size of the cursor boxes in the Persist image. Changing the
size will not change the currently displayed image. The MSIZ? query returns the current
settings of MSIZ based on the current setting of the system level USIZ command.
Command syntax
MSIZ
width
height
or
MSIZ
common_size
size
positive floating point number
common_size
square box
Query syntax
MSIZ?
Query response
width
,
height
Other required
commands
The correct unit of measure should be selected with the
USIZ
command prior to setting
the size.
Example
USIZ 1
// Select inches for units
MSIZ 2.0
// Set size to 2.0 inches
IMGL BriteBox // Select BriteBox image
IMGU
// Draw selected image using new size
Summary of Contents for 802BT
Page 1: ...802BT 802R Video Test Generator User Guide ...
Page 52: ...42 Chapter 2 Testing Video Displays ...
Page 62: ...52 Chapter 3 Administrative Tasks ...
Page 118: ...108 Chapter 7 Using GPIB Interface ...
Page 194: ...184 Chapter 10 Testing EDID ...
Page 204: ...194 Chapter 11 Testing HDCP ...
Page 214: ...204 Appendix A Command Reference Commands by name ...
Page 595: ...Video Test Generator User Guide 585 ...
Page 611: ...Video Test Generator User Guide 601 ...
Page 612: ...602 Appendix A Command Reference ...
Page 716: ...706 Appendix B Image Reference ...
Page 744: ...734 Appendix C Error Messages ...