Video Test Generator User Guide
563
VSIZ
Vertical SIZe
Class
Format parameter setting - Video resolution
Description
Establishes the vertical physical size of the image on the display. Units expected (or
returned) vary according to the last mode set with
USIZ
command. The VSIZ command is
context sensitive and must appear between begin and end commands
FMTB
and
FMTE
.
The VSIZ? query returns the current setting of VSIZ.
Note: Make sure that the USIZ parameter is properly set before using the VSIZ command.
Changing the USIZ setting after entering VSIZ will convert the size to match the new unit
of measure.
Command syntax
VSIZ
size
size
physical size; positive value (floating point accepted)
Query syntax
VSIZ?
Query response
size
Other required
commands
The units of measure must be properly set by
USIZ
before entering VSIZ. The
ALLU
command updates hardware to the new setting and redraws the test image, taking the
new size into account.
Example
FMTB
// Begin editing session
// One or more format editing commands ...
USIZ 1
// Select inches as unit of measure in buffer
HSIZ 10.4 // Set width to 10.4 in buffer
VSIZ 7.8
// Set height to 7.8 in buffer
ALLU
// Test the new settings
// One or more format editing commands ...
FMTE
// End editing session
Summary of Contents for 802BT
Page 1: ...802BT 802R Video Test Generator User Guide ...
Page 52: ...42 Chapter 2 Testing Video Displays ...
Page 62: ...52 Chapter 3 Administrative Tasks ...
Page 118: ...108 Chapter 7 Using GPIB Interface ...
Page 194: ...184 Chapter 10 Testing EDID ...
Page 204: ...194 Chapter 11 Testing HDCP ...
Page 214: ...204 Appendix A Command Reference Commands by name ...
Page 595: ...Video Test Generator User Guide 585 ...
Page 611: ...Video Test Generator User Guide 601 ...
Page 612: ...602 Appendix A Command Reference ...
Page 716: ...706 Appendix B Image Reference ...
Page 744: ...734 Appendix C Error Messages ...