Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
Planmeca Proline CC panoramic x-ray E-27
Technical Manual
4
PATIENT POSITIONING MECHANISM
4.1
Checking the mechanism
You should take an exposure of the ball phantom to check that the patient positioning mecha-
nism is correctly positioned.
Figure
50
Remove the sensor alignment tool from the adapter mechanism and attach the sensor head to its
position (for fixed sensor head refer to section “Removing and attaching the fixed sensor head”
on page E-4).
Slide the ball phantom into the patient positioning mechanism.
Adjust the position of the patient positioning mechanism to 00.
Enter the service mode.
Select a kilovolt value of 60 and a milliampere value of 2. Note that it is only possible to select
this low mA-value in the service mode.
Exit the service mode.
Drive the x-ray unit to the exposure position. The rotating unit will turn to the start position and the
indicator light will come on.
NOTE
The ready indicator light will be flashing if the PC is not in the ready state. Refer to the
Dimaxis user’s manual on how to prepare the system ready for the exposure.
Take an exposure using the normal resolution (refer to the Dimaxis user’s manual). Hold the
exposure down for the duration of the exposure.
CAUTION
Radiation is generated when exposure button is pressed.
START
END
PLANMECA
No.50971
X-LINE
Y-LIN
E
Ball phantom
°
mm
0 0
kV
mA
2
6 0
READY
PRET
Summary of Contents for Proline CC
Page 2: ......
Page 42: ...Chapter C HELP ERROR MESSAGES C 12 Planmeca Proline CC panoramic x ray Technical Manual...
Page 186: ...Chapter H PARTS REPLACEMENT REPAIR H 8 Planmeca Proline CC panoramic x ray Technical Manual...
Page 202: ...Chapter I FUNCTIONAL DESCRIPTION I 16 Planmeca Proline CC panoramic x ray Technical Manual...
Page 203: ...Planmeca Proline CC panoramic x ray J 1 Chapter Technical Manual J SCHEMATICS DIAGRAMS...
Page 204: ...Chapter J SCHEMATICS DIAGRAMS J 2 Planmeca Proline CC panoramic x ray Technical Manual...
Page 206: ......