Chapter
D - ADJUSTMENT & CALIBRATION
D-4 Planmeca Proline CC panoramic x-ray
Technical Manual
2
PANORAMIC X-RAY BEAM ADJUSTMENT
NOTE
In case the x-ray unit is equipped with optional AEC (Automatic Exposure Control), the
AEC calibration must be performed after the panoramic x-ray beam adjustment. See
instructions given in section “AUTOMATIC EXPOSURE CONTROL (AEC) CALIBRATION”
on page D-26
2.1
Checking the panoramic beam position
Before the x-ray is used the position of the radiation beam must be checked. Remove the inner
covers from tube head assembly and the cassette drive assembly. Four screws hold the tube
head cover in position, these are located at the rear of the tube head assembly. The screws that
hold the cassette drive cover in place are located at the front and lower side of the cover. The pri-
mary slot mechanism is located under the tube head cover.
Figure
1
Switch the unit on. Enter the test mode (see section “Test mode” on page B-4).
Press the ready key to drive the x-ray into the exposure position. The rotating unit and cassette
carriage will move to their respective ready positions. When the x-ray is ready the indicator light
will come on.
The horizontal sector display will remain empty which indicates that the primary slot number 0
has been selected.
Figure
2
Start to take a test exposure by pressing and holding down the exposure button on the remote
control. Stop pressing the exposure button when the tube head has rotated to a convenient posi-
tion for viewing the secondary slot.
An error code will appear on the display and start to flash.
Tube head cover
Cassette drive cover
READY
PRET
Primary slot number 0
E r 0 0
Summary of Contents for Proline CC
Page 2: ......
Page 42: ...Chapter C HELP ERROR MESSAGES C 12 Planmeca Proline CC panoramic x ray Technical Manual...
Page 186: ...Chapter H PARTS REPLACEMENT REPAIR H 8 Planmeca Proline CC panoramic x ray Technical Manual...
Page 202: ...Chapter I FUNCTIONAL DESCRIPTION I 16 Planmeca Proline CC panoramic x ray Technical Manual...
Page 203: ...Planmeca Proline CC panoramic x ray J 1 Chapter Technical Manual J SCHEMATICS DIAGRAMS...
Page 204: ...Chapter J SCHEMATICS DIAGRAMS J 2 Planmeca Proline CC panoramic x ray Technical Manual...
Page 206: ......