Chapter
D - ADJUSTMENT & CALIBRATION
D-34 Planmeca Proline CC panoramic x-ray
Technical Manual
9.2
Checking the head support position
Press the ceph mode selection key to enter the ceph mode if you are not already in it. The rotat-
ing unit and cassette carriage will move automatically to the correct position for taking cephalom-
etric exposures. Collimator number 4 will be automatically selected.
Remove the tube head covers and cassette carriage covers (see sections “Tube head covers” on
page H-1, “Secondary slot cover” on page H-1 and “Cassette carriage back cover” on page H-1).
Place the ceph head support alignment tools in the ear post holders as far as they go (Fig. 36, 1).
Rotate the head support to the 90° position.
Look through the opening on the cephalostat cassette holder and the opening on the head sup-
port alignment tools. The fourth primary slot is seen symmetrically if the adjustment is correct
(Fig. 36, 2).
Figure
36
If the fourth primary slot is not seen symmetrically, the head support requires readjustment. See
instruction given in section “Head support adjustment” on page D-35.
4
CEPH
1
2
Summary of Contents for Proline CC
Page 2: ......
Page 42: ...Chapter C HELP ERROR MESSAGES C 12 Planmeca Proline CC panoramic x ray Technical Manual...
Page 186: ...Chapter H PARTS REPLACEMENT REPAIR H 8 Planmeca Proline CC panoramic x ray Technical Manual...
Page 202: ...Chapter I FUNCTIONAL DESCRIPTION I 16 Planmeca Proline CC panoramic x ray Technical Manual...
Page 203: ...Planmeca Proline CC panoramic x ray J 1 Chapter Technical Manual J SCHEMATICS DIAGRAMS...
Page 204: ...Chapter J SCHEMATICS DIAGRAMS J 2 Planmeca Proline CC panoramic x ray Technical Manual...
Page 206: ......