Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
Planmeca Proline CC panoramic x-ray E-49
Technical Manual
The radiation beam must appear on the beam alignment tool as shown on the figure below. It
must reach, but not overlap,
the middle line of the tool. If it does not, the cephalostat arm posi-
tion must be adjusted. Note that the beam does not have to reach the right side line. Refer to the
section “Adjusting the cephalostat arm position” on page E-51.
Figure
83
Press the primary selection key once. The letter “d” will appear on the display.
Stand behind the tube head and protect yourself from radiation. Press and hold down the expo-
sure button. The image of the radiation beam will appear on the beam alignment tool.
The radiation beam must appear symmetrically on the beam alignment tool as shown on the fig-
ure below. If it does not refer to the section “Radiation beam adjustment” on page E-14. The
adjustment is performed in the same way as when adjusting the standard panoramic slot.
Figure
84
D2alinmhor
iz.eps
Right side line
Middle line
d
D2adj_2pr
im.eps
A
A
Summary of Contents for Proline CC
Page 2: ......
Page 42: ...Chapter C HELP ERROR MESSAGES C 12 Planmeca Proline CC panoramic x ray Technical Manual...
Page 186: ...Chapter H PARTS REPLACEMENT REPAIR H 8 Planmeca Proline CC panoramic x ray Technical Manual...
Page 202: ...Chapter I FUNCTIONAL DESCRIPTION I 16 Planmeca Proline CC panoramic x ray Technical Manual...
Page 203: ...Planmeca Proline CC panoramic x ray J 1 Chapter Technical Manual J SCHEMATICS DIAGRAMS...
Page 204: ...Chapter J SCHEMATICS DIAGRAMS J 2 Planmeca Proline CC panoramic x ray Technical Manual...
Page 206: ......