Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
Planmeca Proline CC panoramic x-ray E-47
Technical Manual
Adjust the head support angle with the two screws located on the left side of the mechanism (use
2.5mm allen key) (see Fig. 81) so that the sensor alignment tool is perpendicular to the cephalo-
stat head support.
Figure
81
Spirit level