Chapter
D - ADJUSTMENT & CALIBRATION
D-42 Planmeca Proline CC panoramic x-ray
Technical Manual
Turn the mirror bracket (Fig. 48, 1) until the light beam is in right position.
Figure
48
Turn the bulb housing (Fig. 49, 1) until the light beam is vertical. Fasten the screws of the bulb
fixing plate (Fig. 49, 2).
Figure
49
4
5
6
PLANMECA PM 2002 Proline CC Cephalostat 18x24
No.652232
Posiotioning light 0mm
1
4
5
6
PLANMECA PM 2002 Proline CC Cephalostat 18x24
No.652232
Posiotioning light 0mm
1
2