Planmeca Proline CC panoramic x-ray F-1
Chapter
Technical Manual
F
PREVENTIVE MAINTENANCE
1
SYSTEM MAINTENANCE
1.1
Cleaning
CAUTION
When disinfecting the unit surfaces, always disconnect the unit from the mains. The unit
must not be exposed to gaseous disinfectants or explosive anesthetics. Never spill any
liquids into the unit.
The chin support and the Ceph cassette holder can be cleaned with NON-ALCOHOLIC disinfec-
tion solutions.
The bite piece of the chin rest can be removed and cleaned with alcohol-based solutions. The
patient support handles can be cleaned with alcohol-based solutions.
Other unit surfaces can be cleaned with a soft cloth damped in a mild cleaning solution.
1.2
Operator’s checks
Exposure indicators
Confirm that the exposure indicator lights turn on in the control panel and in the remote control
for the length of the exposure.
Exposure warning signal
Confirm that the buzzer comes on for the length of the exposure. If needed, adjust the tone of the
signal, see section “EXPOSURE WARNING SIGNAL ADJUSTMENT” on page D-25.
Exposure switch
Confirm that the exposure switch requires continuous activation to maintain the exposure.
Releasing the exposure switch during the radiation should produce the ER00 message on the
displays (clear it by pressing the CTL-key once). Make a visual check that the exposure switch
spiral cable is not damaged.
Labels
Check that no labels are detached or worn and that they are all legible.
Summary of Contents for Proline CC
Page 2: ......
Page 42: ...Chapter C HELP ERROR MESSAGES C 12 Planmeca Proline CC panoramic x ray Technical Manual...
Page 186: ...Chapter H PARTS REPLACEMENT REPAIR H 8 Planmeca Proline CC panoramic x ray Technical Manual...
Page 202: ...Chapter I FUNCTIONAL DESCRIPTION I 16 Planmeca Proline CC panoramic x ray Technical Manual...
Page 203: ...Planmeca Proline CC panoramic x ray J 1 Chapter Technical Manual J SCHEMATICS DIAGRAMS...
Page 204: ...Chapter J SCHEMATICS DIAGRAMS J 2 Planmeca Proline CC panoramic x ray Technical Manual...
Page 206: ......