Chapter
D - ADJUSTMENT & CALIBRATION
D-38 Planmeca Proline CC panoramic x-ray
Technical Manual
Slide the soft tissue filter positioning light to left as far as possible. The light will come on and the
soft tissue filter, on the cassette carriage, will automatically move to the correct position.
Figure
41
The beam from the soft tissue filter positioning light must strike the line on the beam alignment
tool marked “Positioning light 0mm” and the number on patient positioning display must be 00. If
the light is in the wrong position it must be repositioned and if the display number is wrong it must
be recalibrated. If the light beam itself is crooked or out of focus the lens or mirror must be repo-
sitioned. Refer to the section “Soft tissue filter positioning light calibration and adjustment” on
page D-40.
Figure
42
Select kilovolt and milliampere values high enough to enable the radiation beam to be seen in a
darkened room for example 62kV and 6mA. The actual values will depend on how dark the room
is.
Stand behind the tube head and protect yourself from radiation. Press and hold down the expo-
sure button. The image of the radiation beam from slot number 4 will appear on the alignment
tool.
Patient positioning light
Beam alignment tool
4
5
6
PLANMECA PM 2002 Proline Cephalostat 18x24 No. 652232
Soft tissue filter 0mm
Positioning light 0mm
Positioning light 25mm
°
mm
“Positioning light 0mm” line
Light beam position
0 0
Summary of Contents for Proline CC
Page 2: ......
Page 42: ...Chapter C HELP ERROR MESSAGES C 12 Planmeca Proline CC panoramic x ray Technical Manual...
Page 186: ...Chapter H PARTS REPLACEMENT REPAIR H 8 Planmeca Proline CC panoramic x ray Technical Manual...
Page 202: ...Chapter I FUNCTIONAL DESCRIPTION I 16 Planmeca Proline CC panoramic x ray Technical Manual...
Page 203: ...Planmeca Proline CC panoramic x ray J 1 Chapter Technical Manual J SCHEMATICS DIAGRAMS...
Page 204: ...Chapter J SCHEMATICS DIAGRAMS J 2 Planmeca Proline CC panoramic x ray Technical Manual...
Page 206: ......