Chapter
D - ADJUSTMENT & CALIBRATION
Planmeca Proline CC panoramic x-ray D-17
Technical Manual
Leave the ball phantom tool in position and place the alignment ruler in position between the pri-
mary slot and the secondary slot as described earlier. The line on the alignment ruler must line
up with the x-line on the ball phantom (Fig. 25).
Figure
25
0° position
Enter the service mode.
Press either of the patient positioning keys once to switch the patient positioning lights on then a
second time to adjust the position of the patient positioning mechanism until the line on the ball
phantom lines up with the line on the alignment ruler. The position of the mechanism will appear
on the display.
Press the CTL-key. The indicator light will come on.
Hold the patient positioning plus (+) key down (about five seconds) until you hear a signal and
the millimeter display zeroes. This indicates that the new position of the patient positioning mech-
anism has been programmed into the memory.
Remove the alignment pin and exit the service mode.
Take another ball phantom picture to check the alignment again. Readjust if necessary.
Checking the eccentricity of the rotation movement
The eccentricity of the rotation movement must be checked after the patient positioning mecha-
nism adjustment. This is done by comparing the distances between the line on the alignment
ruler and the x-line on the ball phantom in “0° position” and in “180° position”. If the deviation is
more than ±1 mm, the position of the secondary slot must be adjusted.
Entering the 180° position:
Enter the test mode.
Press the ready key to drive the x-ray unit to the exposure position. When the unit is ready the
indicator light will come on.
Start to take a test exposure by pressing and holding down exposure button on the remote con-
trol. Stop pressing the exposure button after about 16 seconds so that the rotating unit stops
moving (near the end position).
An error code will appear on the display. Clear the error code with the CTL-key.
START
END
PLANMECA
No.50971
X-LINE
Y-LINE
X-LINECA.EPS
Alignment ruler
Ball phantom
Tube head side
Secondary slot side
°
mm
1 2
CTL
°
mm
0 0
READY
PRET
E r 0 0
Summary of Contents for Proline CC
Page 2: ......
Page 42: ...Chapter C HELP ERROR MESSAGES C 12 Planmeca Proline CC panoramic x ray Technical Manual...
Page 186: ...Chapter H PARTS REPLACEMENT REPAIR H 8 Planmeca Proline CC panoramic x ray Technical Manual...
Page 202: ...Chapter I FUNCTIONAL DESCRIPTION I 16 Planmeca Proline CC panoramic x ray Technical Manual...
Page 203: ...Planmeca Proline CC panoramic x ray J 1 Chapter Technical Manual J SCHEMATICS DIAGRAMS...
Page 204: ...Chapter J SCHEMATICS DIAGRAMS J 2 Planmeca Proline CC panoramic x ray Technical Manual...
Page 206: ......