Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
E-18 Planmeca Proline CC panoramic x-ray
Technical Manual
NOTE
Always use the plus (+) key for the final adjustment. This makes sure that the slot mecha-
nism is correctly positioned by taking up the play between the draw nut and the position-
ing thread.
For example, if you are adjusting the beam to the left by pressing the minus (-) key you should
move the slot mechanism a little further to the left than necessary, and then press the plus (+)
key to move it back to the right to the correct position.
Figure
24
If you cannot adjust the beam far enough to the left using the method described above you may
have to move the metal tab on the draw nut. Loosen the securing screw rotate the tab slightly
and then tighten the screw. The slot position should then be rechecked.
Figure
25
Press the horizontal segment selection key to store the slot position in the memory and move on
to the next slot. The number of the next slot will appear on the display and start to flash.
Repeat the alignment procedure described above for this slot and any other slots.
Exit the primary slot calibration mode and check the position of all the beams again. Realign if
necessary.
Exit the service mode.
Final adjustment
Position thread
Draw nut
in this direction
Metal tab
Positioning thread
Draw nut
2
CTL
Summary of Contents for Proline CC
Page 2: ......
Page 42: ...Chapter C HELP ERROR MESSAGES C 12 Planmeca Proline CC panoramic x ray Technical Manual...
Page 186: ...Chapter H PARTS REPLACEMENT REPAIR H 8 Planmeca Proline CC panoramic x ray Technical Manual...
Page 202: ...Chapter I FUNCTIONAL DESCRIPTION I 16 Planmeca Proline CC panoramic x ray Technical Manual...
Page 203: ...Planmeca Proline CC panoramic x ray J 1 Chapter Technical Manual J SCHEMATICS DIAGRAMS...
Page 204: ...Chapter J SCHEMATICS DIAGRAMS J 2 Planmeca Proline CC panoramic x ray Technical Manual...
Page 206: ......