Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
Planmeca Proline CC panoramic x-ray E-25
Technical Manual
Taking a ball phantom exposure
You can now check the adjustments by taking a ball phantom exposure either as described in
section “Checking the mechanism” on page E-27 or using the Dimax2Tool program’s Tests / Ball
Phantom command.
Slide the ball phantom into the patient positioning mechanism. Position the layer light beam so
that it is on the black reference line on the side of the ball phantom. Select the standard pano-
ramic exposure mode and Select a kilovolt value of 66 and a milliampere value of 2.
NOTE
If the temple supports are in position, remove them before taking the ball phantom expo-
sure.
Select Ball Phantom from Tests menu.
Figure
46
The pmpcal2 window with text Waiting appears.
Figure
47
Press the ready key to drive the x-ray to the exposure position.
The text Exposure appears.
Figure
48