Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
Planmeca Proline CC panoramic x-ray E-17
Technical Manual
Primary slot adjustment - radiation beam too far to the right or left
WARNING
WARNING: DO NOT ADJUST THE POSITION OF THE PRIMARY SLOT MECHANISM WHILE
X-RAYS ARE BEING GENERATED. CHECK THE BEAM POSITION, ADJUST THE POSITION
WITH THE X-RAYS SWITCHED OFF AND THEN RECHECK THE POSITION. IF THE BEAM
IS STILL MISALIGNED REPEAT THE OPERATION.
If any of the beams are too far to the left or right of the alignment rectangle they must be cen-
tered.
Figure
23
NOTE
A microprocessor controlled motor determines the position of each slot. If the beam is
readjusted to the left or right the new position must be reprogrammed into the memory.
Enter the service mode if you are not already in it.
Press the CTL-key.
Press and hold down the primary slot key until the number of the slot that you are checking
appears on the primary slot display and starts to flash. You are now in the primary slot calibration
mode.
For example, if you have just checked the position of the beam from the slot the number 1 will
appear on the display.
Adjust the position of the beam by pressing the patient positioning (mm) keys. The minus key will
move the slot mechanism, and beam to the left, and the plus key will move the slot mechanism,
and beam to the right.
Check the position of the beam again and, if necessary, repeat the above alignment procedure
until the beam is within the alignment rectangle on the fluorescent screen.
CAUTION
Radiation is generated when the exposure button is pressed. Take adequate protection
measures. Keep the exposure time as short as possible.
°
mm
hold for 4 seconds
CTL
CTL
UP
DOWN
PM 2002 CC X-RAY BEAM ALIGNMENT SCREEN no. 50972
NOTE: DO NOT LEAVE SCREEN IN DAYLIGHT
UP
DOWN
PM 2002 CC X-RAY BEAM ALIGNMENT SCREEN no. 50972
NOTE: DO NOT LEAVE SCREEN IN DAYLIGHT
Beam too far to the right
Beam too far to the left
CTL
1
Summary of Contents for Proline CC
Page 2: ......
Page 42: ...Chapter C HELP ERROR MESSAGES C 12 Planmeca Proline CC panoramic x ray Technical Manual...
Page 186: ...Chapter H PARTS REPLACEMENT REPAIR H 8 Planmeca Proline CC panoramic x ray Technical Manual...
Page 202: ...Chapter I FUNCTIONAL DESCRIPTION I 16 Planmeca Proline CC panoramic x ray Technical Manual...
Page 203: ...Planmeca Proline CC panoramic x ray J 1 Chapter Technical Manual J SCHEMATICS DIAGRAMS...
Page 204: ...Chapter J SCHEMATICS DIAGRAMS J 2 Planmeca Proline CC panoramic x ray Technical Manual...
Page 206: ......