D - ADJUSTMENT & CALIBRATION
Planmeca Proline CC panoramic x-ray D-19
Technical Manual
4
PATIENT POSITIONING LIGHTS
4.1
Checking the lights
The positions of the light beams from the three patient positioning lights should be checked.
Place the Frankfort plane alignment tool into position on top of the ball phantom.
Figure
27
Press either of the patient positioning keys to switch the positioning lights on and then a second
time to drive the patient positioning mechanism to 00 if it is not already there.
The Frankfort plane light beam should be located between the two black lines on the alignment
tool. It should be horizontal.
Figure
28
The plane-in-focus light beam should be positioned so that it is on the black reference line on the
side of the ball phantom.
Figure
29
START
END
PLANMECA
No.50971
X-LINE
Y-LINE
PLANMECA OY
NO.50977
Frankfort plane alignment tool
Ball phantom
°
mm
0 0
STA
RT
END
PLANM
EC
A
No.50971
X-LINE
Y-LIN
E
PLANMECA OY
NO
.50977
Frankfort plane light beam
START
END
PLANMECA
No.50971
X-LINE
Y-LINE
PL
AN
ME
CA
O
Y
NO
.50
977
Plane-in-focus light beam
Summary of Contents for Proline CC
Page 2: ......
Page 42: ...Chapter C HELP ERROR MESSAGES C 12 Planmeca Proline CC panoramic x ray Technical Manual...
Page 186: ...Chapter H PARTS REPLACEMENT REPAIR H 8 Planmeca Proline CC panoramic x ray Technical Manual...
Page 202: ...Chapter I FUNCTIONAL DESCRIPTION I 16 Planmeca Proline CC panoramic x ray Technical Manual...
Page 203: ...Planmeca Proline CC panoramic x ray J 1 Chapter Technical Manual J SCHEMATICS DIAGRAMS...
Page 204: ...Chapter J SCHEMATICS DIAGRAMS J 2 Planmeca Proline CC panoramic x ray Technical Manual...
Page 206: ......