Chapter
D - ADJUSTMENT & CALIBRATION
D-40 Planmeca Proline CC panoramic x-ray
Technical Manual
9.5
Soft tissue filter positioning light calibration and adjustment
Soft tissue filter positioning light calibration
Enter the service mode (see section “How to enter/exit the service mode” on page B-6).
Orientate the Cephalometric beam alignment tool so that it is vertical and then place it in the
cephalometric cassette holder. Position it so that it is on the left of the cassette holder.
Figure
45
Press the CTL-key, the light beam position number will appear on the main display.
Slide the light to the right as far as possible and check that the position number on the display is
>10. If it is not, adjust the light assembly position with the light assembly lever (Fig. 46, 1) so that
the number will be >10.
Figure
46
Press the Vertical sector selection D key. The text “CAL” will appear on the main display and the
rotating unit and cassette carriage will move automatically to the cephalometric mode position.
You are now in the soft tissue filter positioning light calibration mode.
Slide the patient positioning light to left as far as possible. The beam from the soft tissue filter
positioning light should strike the line on the beam alignment tool marked “Positioning light 0mm”
(Fig. 46, 2).
4
5
6
PLANMECA PM 2002 Proline Cephalostat 18x24 No. 652232
Soft tissue filter 0mm
Positioning light 0mm
Positioning light 25mm
Cassette
holder
Beam
alignment
tool
4
5
6
PLANMECA PM 2002 Proline CC Cephalostat 18x24
No.652232
Soft tissue filter 0mm
Posiotioning light 25mm
Posiotioning light 0mm
1
3
4
2
0 - 48mm
Summary of Contents for Proline CC
Page 2: ......
Page 42: ...Chapter C HELP ERROR MESSAGES C 12 Planmeca Proline CC panoramic x ray Technical Manual...
Page 186: ...Chapter H PARTS REPLACEMENT REPAIR H 8 Planmeca Proline CC panoramic x ray Technical Manual...
Page 202: ...Chapter I FUNCTIONAL DESCRIPTION I 16 Planmeca Proline CC panoramic x ray Technical Manual...
Page 203: ...Planmeca Proline CC panoramic x ray J 1 Chapter Technical Manual J SCHEMATICS DIAGRAMS...
Page 204: ...Chapter J SCHEMATICS DIAGRAMS J 2 Planmeca Proline CC panoramic x ray Technical Manual...
Page 206: ......